How do you calibrate a micropipette?

image

Micropipette calibration is typically performed using the gravimetric method, where the volume dispensed is determined by measuring the weight of distilled water.

๐’๐ญ๐ž๐ฉ-๐›๐ฒ-๐’๐ญ๐ž๐ฉ ๐†๐ซ๐š๐ฏ๐ข๐ฆ๐ž๐ญ๐ซ๐ข๐œ ๐‚๐š๐ฅ๐ข๐›๐ซ๐š๐ญ๐ข๐จ๐ง

  • ๐€๐œ๐œ๐ฅ๐ข๐ฆ๐š๐ญ๐ž ๐„๐ช๐ฎ๐ข๐ฉ๐ฆ๐ž๐ง๐ญ: Place the micropipette, distilled water, and tips in the calibration room for at least 2 hours to reach a stable temperature (ideally between 20ยฐC and 25ยฐC)
  • ๐’๐ž๐ญ๐ฎ๐ฉ ๐ญ๐ก๐ž ๐๐š๐ฅ๐š๐ง๐œ๐ž: Place a small, clean beaker containing 3 mm of distilled water onto an analytical balance and zero the scale.
  • ๐๐ซ๐ž-๐ฐ๐ž๐ญ ๐ญ๐ก๐ž ๐“๐ข๐ฉ: Aspirate and dispense the target volume 3โ€“5 times to humidify the air inside the tip and pipette.
  • ๐–๐ž๐ข๐ ๐ก ๐ญ๐ก๐ž ๐•๐จ๐ฅ๐ฎ๐ฆ๐ž: : Aspirate the liquid, touch the tip to the inner wall of the beaker at a 45ยฐ angle, and dispense the full amount. Record the weight, re-zero the balance, and repeat this at least 10 times at both the minimum and maximum volumes of the pipette.
  • ๐‚๐š๐ฅ๐œ๐ฎ๐ฅ๐š๐ญ๐ž ๐‘๐ž๐ฌ๐ฎ๐ฅ๐ญ๐ฌ: Convert the recorded weight (mg) to volume (ฮผL) by dividing the weight by the density of water (Z-factor) at the recorded temperature.